Thermogravimetric analysis (TGA) loss on drying testing is a high-precision technique that measures the mass loss of a sample as it is heated under a controlled atmosphere, quantifying moisture, residual solvents, and other volatile components within minutes. It is widely applied to polymers, pharmaceuticals, foods, chemicals, and advanced materials for quality control, incoming material inspection, and research and development. This article explains the working principle, applicable standards, sample requirements, testing procedures, and typical applications of TGA-based loss on drying experiments.
Technical Overview
Thermogravimetric analysis (TGA) is an analytical technique that continuously records the mass of a specimen as a function of temperature or time while the specimen is subjected to a precisely controlled temperature program and purge atmosphere. A thermobalance consists of a sensitive microbalance, typically with a resolution of 0.1 to 1 microgram, a programmable furnace, a sample pan made of alumina or platinum, and a gas flow system supplying inert nitrogen or oxidative air and oxygen atmospheres.
In a loss on drying (LOD) experiment by TGA, the sample is heated from ambient temperature to a specified drying temperature, or held isothermally at that temperature, and the percentage of mass lost is recorded. This mass loss corresponds to the release of surface moisture, bound water, residual solvents, and other volatile substances. The resulting thermogravimetric (TG) curve and its derivative (DTG curve) allow overlapping mass-loss events to be separated by their characteristic temperature ranges, which is a decisive advantage over conventional oven-drying methods.
Compared with traditional oven or halogen moisture balance procedures, TGA loss on drying testing offers several benefits: it requires only milligram-scale samples, is fully automated, completes within 20 to 60 minutes, and simultaneously provides additional information such as decomposition onset temperature, volatile fraction, combustible content, and ash residue. These capabilities make TGA a powerful tool in accredited third-party laboratories as well as in industrial quality control and research environments.
Testing Samples
Polymers and Plastics
Thermoplastic resins, pellets, granules, films, and fibers, including polyethylene, polypropylene, PET, polyamide, polycarbonate, and ABS, are routinely tested for moisture before melt processing. Excessive moisture in hygroscopic polymers such as PET and polyamide can cause hydrolytic degradation, silver streaks, and bubbles during injection molding or extrusion.
Pharmaceutical and Nutraceutical Materials
Active pharmaceutical ingredients (APIs), excipients, granules, lyophilized powders, and plant extracts are tested for water content and residual solvents, which directly affect flowability, compaction behavior, chemical stability, and shelf life.
Food and Agricultural Products
Milk powder, starches, flours, proteins, spices, tea, and other food ingredients are evaluated for moisture content, which influences nutritional labeling, processing behavior, microbial stability, and storage performance.
Chemicals, Coatings, and Fine Powders
Pigments, dyes, fillers such as calcium carbonate and talc, catalysts, adhesives, and coating powders are tested to verify drying specifications and volatile organic content.
Battery and Advanced Materials
Cathode and anode active materials, conductive carbons, binder powders, and electrolyte salts require strict moisture control, because trace water can trigger side reactions and degrade electrochemical performance and safety.
Inorganic, Mineral, and Rubber Materials
Ceramic powders, clays, ores, coal, coke, and elastomeric compounds can be characterized for moisture, volatile matter, and filler or carbon black content in a single TGA run.
Testing Items
Typical test items and deliverables for TGA loss on drying experiments include:
- Moisture content and total loss on drying at a specified temperature, for example 105 °C
- Volatile matter and residual solvent content released below a defined temperature threshold
- Decomposition onset temperature and thermal stability assessment
- Stepwise compositional analysis: moisture, volatiles, combustible organic fraction, and inorganic ash residue
- Filler, glass fiber, or carbon black content in polymer and rubber compounds
- Derivative (DTG) peak temperatures and maximum mass-loss rates for kinetic comparison
- Residual mass (ash) after oxidative burn-off in air or oxygen atmosphere
Testing Methods
Sample Preparation
A representative specimen of typically 5 to 20 mg, or up to 50 mg for heterogeneous materials, is homogenized and loaded into an open alumina or platinum pan. Samples may be powders, pellets, films, fibers, or small sections cut from larger parts. Care is taken to minimize moisture exchange with ambient air between sampling and analysis.
Instrument Setup and Calibration
The microbalance is calibrated with certified mass standards, and the furnace temperature is calibrated using Curie-point or melting-point reference materials. High-purity nitrogen is supplied at a flow rate of 20 to 60 mL/min for inert drying and decomposition steps, while compressed air or oxygen is used for oxidative ash determination.
Typical Test Procedure
A common loss on drying program is as follows: equilibrate the sample at ambient temperature or 40 °C, heat at 10 °C/min to the specified drying temperature such as 105 °C, and hold isothermally until the mass change falls below a constant-mass criterion, for example less than 0.05% per 5 minutes. For full compositional analysis, the run continues under nitrogen to approximately 600 °C, after which the atmosphere is switched to air or oxygen and the temperature is raised to 800 to 900 °C to burn off carbon and quantify the inorganic ash residue.
Calculation of Results
Loss on drying is calculated as the percentage of mass lost relative to the initial sample mass: LOD (%) = (m0 − m1) / m0 × 100, where m0 is the initial mass and m1 is the mass at the specified endpoint. The DTG curve is used to confirm that the drying step is fully separated from any decomposition event, ensuring the validity of the result.
Reference Standards
- ASTM E1868 — Standard Test Methods for Loss-On-Drying by Thermogravimetry
- ASTM E1131 — Standard Test Method for Compositional Analysis by Thermogravimetry
- ASTM D6370 — Standard Test Method for Rubber Compositional Analysis by Thermogravimetry (TGA)
- ISO 11358 series — Plastics: Thermogravimetry (TG) of polymers
- ISO 9924 series — Rubber and rubber products: Determination of composition by thermogravimetry
- GB/T 6284 — Determination of loss on drying for chemical products
- GB/T 27761 — General rules for thermogravimetric analysis
Applications
TGA loss on drying testing is applied across a broad range of industrial and scientific scenarios:
- Incoming quality control of hygroscopic resins before injection molding and extrusion to prevent hydrolysis and surface defects
- Moisture verification of battery electrode materials and electrolyte salts to ensure cell safety and performance
- Pharmaceutical batch release testing, stability studies, and drying process validation
- Food ingredient moisture determination for labeling compliance, shelf-life estimation, and process control
- Verification of filler, fiber, and carbon black content in polymer compounds and rubber formulations
- Research and development studies on thermal stability, degradation kinetics, and formulation comparison
- Failure analysis of degraded or contaminated parts through characteristic mass-loss fingerprints
- Support of regulatory submissions and customer specifications through testing at accredited third-party laboratories
FAQ
Question: How is TGA loss on drying different from oven drying or a halogen moisture balance?
TGA uses milligram-scale samples and a programmable temperature and atmosphere profile, completing a typical determination in 20 to 60 minutes. It also separates overlapping mass-loss steps and can extend the analysis to decomposition and ash content in the same run. Oven and halogen methods typically require gram-scale samples, fixed temperatures, and longer times, and cannot distinguish water from other volatiles.
Question: Does loss on drying equal water content?
Not necessarily. The TGA mass loss includes all substances volatilized below the specified endpoint, such as residual solvents and low-boiling additives, in addition to water. If a specific water value is required, Karl Fischer titration is recommended as a complementary method.
Question: How do I choose the drying temperature?
The temperature is defined by the relevant product specification or standard. A temperature of 105 °C is common for free water, while some materials require 120 to 130 °C or material-specific conditions. The chosen temperature must remain safely below the decomposition onset to avoid overestimating volatile content.
Question: What sample size and accuracy can be expected?
Typical specimen masses are 5 to 20 mg. With microbalance resolution at the sub-microgram level, mass losses on the order of 0.01% to 0.1% can be reliably detected, depending on sample size and homogeneity.
Question: What is reported in a TGA loss on drying test report?
A professional report includes the initial sample mass, the complete temperature program and atmosphere, the TG and DTG curves, the calculated loss on drying percentage, any additional stepwise mass losses, the residual or ash mass, and the applicable test standard. This documentation ensures traceability and supports comparison with specification limits.